Resources - Technical Notes

Synergy™ H1 Shaking System - Maximum Shaking Amplitude Based on Assay Volume and Plate Type

Download

Related Products: Synergy H1

October 26, 2010

 

Synergy H1 offers broad range of shaking speeds and amplitudes.  If the wells of a microplate are almost full, shaking can result in spillage inside the instrument. The table below is designed to help prevent this. Select your microplate type, assay volume, and you will get the acceptable shake amplitude on your instrument.

Sample volume

Linear

Orbital slow

Orbital fast

Double orbital slow

Double orbital fast

6-well microplate

0 to 3 mL

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

3 to 4 mL

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 mm

4 to 5 mL

1 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 mm

5 to 7 mL

1 mm

1 to 6 mm

1 to 3 mm

1 mm

No

7 to 8 mL

1 mm

No

1 to 3 mm

No

No

9 mL

1 mm

No

No

No

No

12-well microplate

0 to 1.5 mL

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1.5 –2.5 mL

1 to 2 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 mm

2.5 to 3 mL

1 to 2 mm

1 to 6 mm

No

No

1 mm

3 to 4 mL

No

1 to 6 mm

No

No

No

24-well microplate

0 to 0.75 mL

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

0.75 to 1 mL

1 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 1.5 mL

1 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

No

1.5 to 2 mL

1 mm

1 to 6 mm

No

1 to 6 mm

No

48-well microplate

0 to 0.5 mL

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

0.5 to 1 mL

1 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 1.3 mL

No

1 to 6 mm

1 to 6 mm

1 to 6 mm

No

96-well microplate

0 to 0.25 mL

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

1 to 6 mm

 

TN102610_07