Optical Clearing for Improved Confocal Imaging of Thick Specimens

Download

Related Products: Cytation C10

March 23, 2021

Introduction

Developed almost 70 years ago, confocal microscopy has become a mainstay for imaging fluorescently labeled biological samples. Widefield microscopy illuminates biological samples with a diffuse field of light that excites fluorophores above and below the focal plane of interest, resulting in a blurry, less resolved image. By placing a pinhole in a conjugate focal plane between the sample and the light detector (camera or PMT), confocal microscopy blocks contaminating out-of-focus light, thus improving axial (z) resolution (Figure 1). This improved resolution means that finer z-sections can be taken deeper within thick samples, such as spheroids and resected tissue, and three dimensional biological relationships can be better elucidated.

Adhered HT-1080 spheroids.

Figure 1. Adhered HT-1080 spheroids. Adhered 1000-cell HT-1080 spheroids were fixed and stained with Hoechst 33342 and Alexa Fluor 488-phalloidin, then buffer (PBS) was either maintained, or exchanged for ScaleS4 clearing agent or 80% glycerol. Z-stacks were then captured at 20x (0.45NA) in either widefield or confocal (60 μm disc) mode. Maximum intensity projections of the entire z-stack were then generated.


Poor image quality and resolution (e.g. signal-to-noise) can result not only from excited out-of-focus fluorophores, but also from biological matter such as lipids and proteins. In the latter case, in-focus light diffracts as it travels through the sample, reducing the amount of signal that is detected. This diffractive effect is quantified by the refractive index, which is a measure of how the speed of light is influenced as it passes through a material. Matching the refractive index of the mounting medium and the sample with the microscope optics as close as possible will aid in minimizing diffracted light, which in turn will improve the quality of the image.

Poor image quality and resolution (e.g. signal-to-noise) can result not only from excited out-of-focus fluorophores, but also from biological matter such as lipids and proteins. In the latter case, in-focus light diffracts as it travels through the sample, reducing the amount of signal that is detected. This diffractive effect is quantified by the refractive index, which is a measure of how the speed of light is influenced as it passes through a material. Matching the refractive index of the mounting medium and the sample with the microscope optics as close as possible will aid in minimizing diffracted light, which in turn will improve the quality of the image.

Imaging depth limits of adhered HT-1080 spheroids in different imaging media

Figure 2. Imaging depth limits of adhered HT-1080 spheroids in different imaging media. Z-slices from each adhered spheroid in Figure 1 were selected when nuclei become unresolvable for PBS (upper panels), or for ScaleS4 or 80% glycerol (lower panels).


Sample Preparation for Imaging HT1080 Spheroids

HT-1080 spheroids were formed by seeding 1000 cells into ultra-low attachment (ULA) U-bottom 96-well plates, #4520, from Corning (Corning, NY) and allowed to coalesce for 24 hours in Advanced DMEM (Gibco) from Thermo Fisher Scientific (Waltham, MA) at 37 °C. Spheroids were then transferred to clear, flat-bottom 96-well plates, #204626, from Agilent Technologies (Santa Clara, CA) (1 spheroid/well) and allowed to settle and establish adhesion at RT for 1 hour. Plates containing spheroids were then transferred back to the 37 °C incubator and allowed to attach to the culture area and spread overnight. Attached spheroids with migrating cell regions were fixed with 4% PFA for 30 minutes, then permeabilized with 0.5% for 1 hour. To achieve full penetration of dyes, spheroids were stained with Hoechst 34580 and Alexa Fluor 488-phalloidin overnight at 4 °C. Spheroids were then washed, and PBS was either maintained or exchanged for either a homemade mounting medium (80% glycerol, 20 mM tris, pH 8.0, 0.5% n-propyl gallate, 0.05% NaN3) or the ScaleS4 (40% D-sorbitol, 10% glycerol, 4M urea, 15% DMSO, 0.5% n-propyl gallate, 0.05% NaN3) (Figure 1).

Imaging Procedure and Processing

Confocal and widefield z-stack images of attached HT-1080 spheroids were captured at 40x using the Cytation C10 confocal imaging reader and Gen5 software in Manual Mode (IMM). The dynamic range of each channel was maximized by selecting the “show saturated pixels” button. The exposure settings for each channel were adjusted such that they were within range, and there were no saturated pixels. The z-range was determined by toggling below and above the sample until the structure of interest (in this example, a spheroid) faded in intensity. The z-step size was set using Nyquist sampling recommendations; for imaging Alexa Fluor 488 using the 40x air objective (NA = 0.6), steps sizes were determined to be 1 μm for widefield, and 0.6 μm for confocal. Images were then subjected to 5 iterations of deconvolution using a point spread function based on the objective. Deconvolved images were then pre-processed to reduce background signal. The rolling ball size for background reduction for Hoechst 34580 (DAPI channel) and Alexa Fluor 488-phalloidin was set to 25 and 50 μm, respectively. Equivalent z-heights for all three conditions (PBS, 80% glycerol, ScaleS4) was compared to determine imaging depth limits (Figure 2).

Conclusion

Confocal microscopy is a powerful mode of imaging that improves image quality by removing out-of-focus light, however sources of diffracted fluorescent light inherent to the sample prevent this improvement to truly be realized. Use of clearing agents allows matching the refractive index of the sample, in this case HT-1080 spheroids, with the refractive index of the optical system allowing imaging depth and resolution to be improved.


For Research Use Only. Not for use in diagnostic procedures.

 