General
Detection modes | Absorbance |
Read methods | Discontinuous kinetic |
Microplate types | 96-well microplates (with cover seal) |
Microplate capacity | 4-microplate capacity |
Temperature control |
Incubation to 45 °C with Condensation Control™ Variation +0.5 °C at 37 °C Plate to plate uniformity +0.5 °C at 37 °C (measured at plate center) |
Shaking | Orbital, user-selectable velocity |
Software | LogPhase App included; provides reader control, data collection and complete analysis. |
Absorbance
Light source | LED |
Detector | 2 photodiodes (measurement & reference) |
Wavelength range | 560 nm – 640 nm, configuration dependent |
Dynamic range | 0 - 4.0 OD |
Resolution | 0.001 OD |
OD accuracy | 0.000 to 2.000 OD + 1% + 0.010 OD |
OD linearity | 0.000 to 2.000 OD + 1% + 0.010 OD |
OD repeatability | 0.000 to 2.000 OD + 1% + 0.005 OD |
Reading speed (kinetic) |
Reading speed: <60 sec per plate Minimum kinetic interval: 2 min 30 sec (<60 sec read time; 90 sec shake) per plate |
Physical Characteristics
Power | 24 VDC power supply compatible with 100-240 volts AC @50-60Hz, 250W (minimum) |
Dimensions | 10.5" H x 26.0" W x 16.0" D (26.7 x 66 x 40.6 cm) |
Weight | 50 lbs (22.7 kg) |
Regulatory
Regulatory | CE and TUV marked. RoHS compliant. |