product design

EL406 Washer Dispenser

Specifications

General

Microplate types 96-, 384- and 1536-well
Low profile and standard height
Solid and filter bottom (option)
Onboard software Create, edit or run multiple protocols
Software (computer control) LHC2 Software
LHC2 Secure for 21 CFR Part 11 compliance (option)
SiLA Compliant driver (option)
Separation Biomagnetic separation, vacuum filtration (optional)
Shaking Programmable up to 60 minutes
Slow, medium, fast or variable
Soaking Programmable up to 60 minutes
Ultrasonic Advantage™ Standard on most configurations
Automation BioStack and 3rd party automation compatible
BioSpa 8 Automated Incubator compatible

Washing

Manifold types 96-well washing: 96-tube manifold - straight dispense tubes
96- and 384-well washing: 96-tube Dual-Action manifold - 20° angled dispense tubes
384-well washing (fast): 192-tube Dual-Action manifold - 7° angled dispense tubes
1536-well washing: Two 32-tube dispense manifolds - 7° angled dispense tubes
Volume range 3 - 3,000 µL/well in 1 µL increments
Wash cycles 1 - 250
Buffer/reagent selection Auto switching module for up to 4 buffers (option)
Supply bottle 4 L or 10 L (optional)
Dispense accuracy +3%
Dispense precision <3% CV (model dependent)
Residual volume <2 µL/well
Wash speed 96 wells, 300 µL/well, 96-tube manifold: 13 seconds
384 wells, 100 µL/well, 192-tube manifold: 17 seconds
1536 wells, 10 µL/wells, two 32-tube manifolds: 36 seconds
Flow rates High flow to low flow
Optimized rates for cell assays
Sterilization Chemical
Vacuum range for filtration Selectable: 0 to -380 mmHg (final at 30s)
Filtration time: 5 - 999 seconds

Dispensing - Peristaltic pump (Multi-Channel)

Manifold types 8-tip (1 x 8) cassette with plastic, 316 SS tubes or sapphire jeweled 316 SS tubes
Fluid delivery 1 or 2 peristaltic pumps
Dispense speed 96 wells, 10 µL /well: 8 seconds
384 wells, 5 µL /well: 12 seconds
1536 wells, 1 µL /well: 27 seconds
Volume range 500 nL - 3,000 µL/well, selectable in 1 µL increments
Flow rates User programmable rates from high to low
Optimized rates for cell assays
Dispense performance 1 µL cassette:
Recommended Volume Range: 1 - 50 µL
Dispense Accuracy: +5% at 1 µL
Dispense Precision: <5% CV at 1 µL
<10% CV at 500 nL
Minimum Prime Volume: 1.20 mL
5 µL cassette:
Recommended Volume Range: 5 - 2,500 µL
Dispense Accuracy: +2.0% at 5 µL
Dispense Precision: <2.5% CV at 5 µL
Minimum Prime Volume: 4.23 mL
10 µL cassette:
Recommended Volume Range: 10 - 3,000 µL
Dispense Accuracy: +2.0% at 10 µL
Dispense Precision: <2.0% CV at 10 µL
Minimum Prime Volume: 7.36 mL
Recommended cassette replacement interval 1 µL Cassette: 1,000 384-well microplates at 5 µL well
5 µL Cassette: 1,000 96-well microplates at 50 µL well
10 µL Cassette: 1,000 96-well microplates at 100 µL well
Sterilization Autoclave, chemical

Dispensing - Syringe pump

Manifold types 96-well dispensing:
One 16-tube (2 x 8) manifold - 316 SS tubes
96-/384-well dispensing:
Two 16-tube (1 x 16) manifolds - 316 SS tubes
1536-well dispensing:
Two 32-tube (1 x 32) manifolds - sapphire jeweled 316 SS or 316 stainless steel tubes
Fluid delivery Two positive displacement syringe drives
Dispense speed 20 µL /well, 96 wells, 1 x 16 tubes: 5.25 seconds
20 µL /well, 384 wells, 1 x 16 tubes: 14 seconds
3 µL /well 1536 wells, 2 x 32 tubes: 7 seconds
Volume range 3 - 3,000 µL/well, selectable in 1 µL increments
Minimum Prime Volume: 12 mL
Flow rates User programmable rates from high to low
Dispense accuracy +1 µL at 5 µL
+1 µL at 20 µL
+1% at 100 µL
Dispense precision <5% CV at 5 µL
<2.5% CV at 20 µL
<1% CV at 100 µL
Supply bottle 1 L or 2 L
Sterilization Chemical, autoclavable option

Physical Characteristics

Power 100 - 240 Volts AC. 50/60 Hz
Dimensions 16.5" W x 18" D x 12.5" H (42 x 46 x 32 cm)
Weight 32 lbs (14.5 kg)

Regulatory

Regulatory CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available.

Specifications are subject to change. Performance values represent the average observed factory test values.

link