General

Detection modes Absorbance
Read methods Endpoint, kinetic, spectral scanning, well area scanning
Microplate types 6- to 384-well plates
Other labware supported Take 3 Micro-Volume Plates
Temperature control not available
Shaking not available
Software Gen5™ Data Analysis Software
Gen5 Secure for 21 CFR Part 11 compliance (option)
Automation BioStack™ and 3rd party automation compatible ('R' configuration)

Absorbance

Light source Xenon flash
Detector photodiode
Wavelength selection monochromator
Wavelength range 200 to 999 nm, in 1 nm increments
Monochromator bandwidth 5 nm
Dynamic range 0 - 4.0 OD
Resolution 0.0001 OD
Pathlength correction yes
Monochromator wavelength accuracy + 2 nm
Monochromator wavelength repeatability + 0.2 nm
OD accuracy 0 to 2 OD: +1% +0.010 OD
2 to 2.5 OD: +3% +0.010 OD
OD linearity 0 to 2 OD: +% +0.010 OD
2 to 2.5 OD: +3% +0.010 OD
OD repeatability 0 to 2 OD: +1% +0.005 OD
2 to 2.5 OD: +3% +0.005 OD
Reading speed (kinetic) 96 wells: 15 seconds
384 wells: 31 seconds

Physical Characteristics

Power 100 - 240 Volts AC. 50/60 Hz.
Dimensions 12" W x 12.5" D x 7.7" H
(30.5 cm x 31.8 cm x 19.6 cm)
Weight <15 lbs (6.8 kg)

Regulatory

Regulatory CE and TUV marked. ROHS compliant. In Vitro Diagnostic use models are available.

Specifications are subject to change. Performance values represent the average observed factory test values.