product design

Gas Controller for CO2 control. Includes CO2 flow meter & tubing set (p/n 1213032)

Part Number: 1210007
Price: $5,814    (Accessory pricing is U.S. list. Local pricing may vary.)

For use with: Synergy H1, Synergy HTX

Gas Controller for CO<sub>2</sub> control. Includes CO2 flow meter & tubing set (p/n 1213032)

Increasingly, life science assays are run on live cells. Some of these assays require control over CO2 and O2 concentrations to modulate the environment for pH buffering or to create a hypoxic condition. These assays are traditionally run in flasks in a low-throughput manner, but there is a burgeoning interest in automation of live cell-based assays in microplates for higher throughput and efficiency, therefore there is a need for cell-friendly microplate instrumentation.

The Gas Controller module for Synergy™ H1  and Synergy HTX allows full control over CO2 concentrations to regulate the environment for microplate based live cell assays. The Gas Controller and other important cell-friendly features of the Synergy H1 and Synergy HTX, like variable shaking parameters, 4-Zone™ temperature control and a robust mechanical system make them ideal microplate readers for cell-based assays.

Synergy H1 with Gas Controller
Synergy H1 with Gas Controller

 



Features

  • Individual or simultaneous control of CO2 and O2 levels
  • CO2 and O2 Control Resolution ± 0.1 %
  • Compact Design
  • Infrared CO2 Sensor
  • Zirconium Oxide O2 Sensor Simple
  • Intuitive control panel
  • Easy to read gas level indicators and alarm LEDs
  • Alarm buzzer on/off


Models

 

1210008     Gas Controller for CO2 and O2 control. For use with Synergy H1 Hybrid Microplate Reader and Synergy HTX Multi-Mode Reader.

1210007     Gas Controller for CO2 control only. For use with Synergy H1 Hybrid Microplate Reader and Synergy HTX Multi-Mode Reader.

1210011     Calibration Kit for CO2 Controller.

01503         Calibrated CO2 Gas Tank



Specifications

 
General
Range
CO2 Control: 0-20%
O2 Control: 1-19%
Control Resolution
CO2 Control: + 0.1%
O2 Control: + 0.1%
Stability
CO2 Control: + 0.2% AT 5% CO2
O2 Control: + 0.2% AT 1% O2
Sensor Type
CO2 Control: Infrared, long lifetime
O2 Control: Zirconium Oxide, long lifetime (no yearly replacement)
Regulatory
Regulatory
All BioTek microplate instrumentation is CE and ETL marked. RoHS Compliant.

Specifications are subject to change