Applications - Technical Notes

Tips for Optimizing Microplate Vacuum Filtration Results For Vacuum-to-Waste Procedures

27-Oct-11

View PDF file

Related Products: EL406, ELx405

 

Author: Wendy Goodrich, Applications Scientist, BioTek Instruments, Inc., Winooski, VT

 

Effective vacuum filtration of assay reagent material from a microplate is a function of many variables including filtration plate design factors such as membrane material, membrane pore size, well size and well shape, and assay parameters, such as the total volume and viscosity of reagents to be removed from the wells. Additionally, test procedure base line settings for vacuum time and pressure may be given assuming only a single vacuum source recommended for the assay, and different vacuum filtration manifold architecture, for example waste tray depth and shape, can influence vacuum draw point patterns across the plate leading to variable results system to system for the same settings. Achieving the goal of even and reproducible distribution of results across the plate is also dependent on the ability to accurately determine when the microplate wells are ‘dry’ (devoid of contaminating or excess substances that can interfere with assay functionality or detection instrument performance thereby effecting final result determination).

This Tech Note offers tips on optimizing vacuum filtration results as applied in DNA sequencing reaction clean-up using BioTek’s ELx405™ vacuum filtration module and Millipore Montage® SEQ96 Sequencing Reaction Clean-up Kit. Although written to accompany the BioTek Application Note, Efficacy of Using a Combination Microplate Washer for Vacuum-Based DNA Sequencing Reaction Clean-up, many of these tips are universal to vacuum filtration using membrane based microplates, and may prove especially useful when a vacuum manifold other than the recommended model is used, when vacuum distribution appears uneven via variable material removal patterns, membrane pore sizes are unknown or unavailable, and time of vacuum is critical to maintain assay performance and sample viability. This Tech Note assumes basic knowledge of using vacuum filtration onboard the BioTek Microplate Washers, and is intended to provide tips for good practice and technique rather than provide instruction for performing a particular assay or instrument function. Differences between the EL406™ and ELx405 vacuum filtration modules, or any other manufacturer model, are not noted and are left to the discrimination of the user.

  1. Establish the recommended vacuum pressure for the assay. An online conversion tool was used to translate the assay recommended 23-25" Hg to 500-600 mmHg units as used by the BioTek ELx405 specifications.
  2. Establish the membrane pore size for the supplied filtration plate. In the referenced Millipore assay this is proprietary information, however even if the exact pore size is unavailable a range of pore size can generally be obtained. For the DNA sequencing clean-up kit the membrane pore size was verified as being less than 0.45 µM (the smallest pore size specified for the BioTek vacuum filtration module).
  3. Using the recommended vacuum pressure and pore size find the closest match using the BioTek selectable vacuum levels, or those provided by the vacuum manifold manufacturer.
  4.  

    Selectable vacuum levels
     
    0.45 µm
    1.2 µm
      96-well 384-well 96-well 384-well
    Lowest - 0.047" 111 mmHg 108 mmHg 117 mmHg 117 mmHg
    Low - 0.032" 200 mmHg 196 mmHg 203 mmHg 198 mmHg
    Medium - 0.020" 405 mmHg 371 mmHg 409 mmHg 387 mmHg
    High - 0 561 mmHg 563 mmHg 550 mmHg 450 mmHg

     

  5. After configuring the manifold for the correct pressure setting, ‘High’ in the case of BioTek’s ELx405™ manifold, identify any other parameters that can be manipulated to optimize for the assay materials. For the DNA sequencing clean-up procedure unspecified membrane pore size would have to be validated as a function of duration or other technique as no further increase in vacuum pressure could be implemented for pore sizes less than 0.45 µm on the ELx405.
  6. Perform a test run using the chosen vacuum pressure at the assay recommended durations using only the wash or other filtration reagent specified for the test. Generally, assay kits provide enough excess reagents for this purpose without the added expense of using real samples or other costly procedures for results determination (capillary electrophoresis in the case of DNA sequencing). Here are some recommendations for the test run:

    a. Weigh the empty filtration plate before loading any reagent.

    b. Add the total amount of reagent to be removed to all wells of the plate (include volume of source material even though it isn’t used for the test run).

    c. Weigh the plate again before filtration.

    d. Perform the vacuum filtration following the manufacturer instructions for time. Even if not required or recommended, blot the underside of the plate during and/or at the end of the cycle. Less blot steps is better, but lower residuals greatly improves assay performance.

    e. Weigh the dry plate following a filtration step. This can provide a good guideline specification for assistance in determining a ‘dry’ plate. For the Millipore procedure a ‘visual’ inspection of the bottom of the wells (not the underside) should result in observation of a ‘shiny’ surface indicative of a dry well. This could also be gauged by a color change or other measure as noted by a given assay.

    f. Adjust time intervals until the fastest vacuum duration within range for the assay produces the lowest residual when the plate is weighed (that weight closest to the dry plate from step 5a). The residual difference could be quite high (>= 1.2 grams for example) and is highly dependent on absorbency rating of membrane material, reagent profile, and duration time (itself depending on factors like total volume in the well and reagent viscosity). A reproducible weight with the lowest window between 5a and 5e, utilizing a duration time closest to the assay specification, provides a satisfactory baseline for successful filtration. Duration sensitivity to recommended filtration times may not be available for an assay, and can be an art as well as a science to determine. Once a ‘dry’ well is confidently determined work backwards, if necessary, decreasing time in 20-45 second intervals until a filtration time within the specified range for the assay is achieved as closely as possible. For the DNA sequencing filtration, this was complicated by the fact that the volume of sample source material was greater than that assumed for the filtration time specified for the assay. Latitude in filtration time was required because there was more total volume to evacuate.

    g. Repeat steps 5b thru 5f for the well volume and reagent required of each filtration cycle. For example, if 20 µL of rinse solution is required for 15 µL of sample, verify filtration time for 35 µL of rinse solution. If the second rinse step requires adding 20 µL of the same or a different solution to the wells verify time for 20 µL of the specified reagent even if it is the same reagent, as determining optimized duration time is highly dependent on well volume in addition to reagent, membrane, and manifold profiles.

  7. When confident removal of material for each volume and filtration time at each step of the assay is achieved, perform a ‘live’ run using a known control for the assay. Some recommendations for the live control run include:

    a. Run enough control wells to give a broad indication of vacuum performance across the plate, but not too much to make the test run overly costly or tax any downstream instrumentation if something fails. Capillary tubes typically used in DNA sequencing runs are very small and sensitive to even minute amounts of contamination or tiny air pockets, so minimizing the impact of any trace reaction artifacts is important not just for sample integrity, but for instrument care. To achieve this, without sacrificing the ability to determine uniform vacuum distribution and thorough washing of the sample during filtration, the control was run on a broad spectrum of wells across the filtration plate, but then transferred to the fewest lanes that could hold all of the control wells on the capillary electrophoresis plate.

    b. Add reagent to any empty wells in the same volume as control or sample wells. The same total well volume should be equal across the plate. The best results are obtained when all wells of the plate contain the same volume and as much of the same reagent material as possible. Wrapping or covering unused wells of the filtration plate to prevent vacuum leak should be avoided as it isn’t a true representation of uniform material removal in the plate.

    c. Use a proven comparative method for test result validation. This assures confident confirmation of test method performance. An added benefit for the DNA sequencing demonstration was that by designing the experiment so that the comparative method strips were run both before and after the control strips allowed confirmation that the test wells did not adversely affect instrument performance.

    d. Analyze results for those variables indicative of method performance. Narrowing down any problems in data improves troubleshooting efficiency. Deviant results from a DNA sequencing run can indicate problems in the sample prep or reaction procedure, not just the clean-up method. For clean-up, a good Length of Read and absence of ‘dye blobs’, or interfering and overlaying peaks of color on the chromatogram, are two characteristic sequencing clean-up success indicators.

  8. Continue to optimize vacuum time and technique using controls until the comparative method and the test method show acceptable correlation. The first control run on the DNA sequencing demonstration resulted in dye blobs clearly indicated on the vacuum filtration wells. This verifies a lack of reaction artifact removal. The pattern of dye blobs was fairly universal across the plate, so an uneven distribution pattern could not be clearly established.

    a. Design follow-up experiments to implement possible solutions specific to each variable requiring optimization. The following conditions were implemented to optimize removal and minimize interference of contaminating material (e.g. excess dye) for DNA sequencing cleanup:

      i. Using the default parameters for clean-up reagent volume, each vacuum filtration step was divided into a ‘long’ filtration, followed by blotting the underside of the filtration plate, then a short ‘dry’ filtration, followed by a final blot. This allowed excess filtrate to be wicked from the membrane after the long filtration thus hastening the final removal of artifact during the short run. The membrane was then also cleaner for the next full filtration cycle that followed.

      ii. The clean-up kit specifications assumed sample volumes of 5 μL and 10 μL, and the sample volume used for the test run was 15 μL. ‘Long’ vacuum time on the first filtration step was increased by 45 seconds to account for the higher sample volume.

      iii. A primer concentration of 5 pmol is recommended for the vacuum filtration method, but the primer concentration of 3.5 pmol employed by the comparative method had been used for the test run. Both primer concentrations were implemented side-by-side on the second run. It is helpful and useful to run original settings in parallel to new settings especially if more than one variable is being optimized.

      iv. The clean-up kit suggests use of a secondary buffer reagent for a rinse step if dye blobs are observed. This was implemented according to the kit insert using default volumes and vacuum times from the first run so as not to introduce additional variables. The buffer was also run at 5 μL higher volume and 45 second longer vacuum duration to match the new rinse solution parameters implemented by 7ii.

      v. Each test condition was run as a full strip to get the broadest possible distribution
      pattern range. This helps in observing even vacuum distribution that may have been
      obscured by the dye blobs on the first run.

  9. Steps employed by 7i showed acceptable results on the second run for the DNA sequencing demonstration requiring no further optimization, but it is highly recommended that if more than one filtration step is required for an assay that the plate be rotated between each step as edge wells are generally more vulnerable to any effects of variable pressure points during applied filtration if they exist.



» View all Tech Resources